SYS 528 Rectangular Sampling Probe
Rectangular isokinetic sampling probe for installed-filter leakage and integrity scanning at a 50 L/min particle-counter flow, designed according to ISO 14644-3 and VDI 2083-3.
Model Overview
| Model | Group | Positioning | Core fit |
|---|---|---|---|
| SYS 528 | Rectangular isokinetic sampling probe | 50 L/min probe for installed-filter leakage and integrity scanning | HEPA/ULPA filter scanning in cleanrooms and clean-air installations |
Key Features
Isokinetic probe geometry
The rectangular inlet is matched to the particle-counter flow and filter downstream velocity to minimise sampling bias.
Smooth internal transition
The rectangular inlet transitions continuously to the round hose outlet, with a smooth internal surface to support particle transport.
Corner-access scanning
The probe shape supports precise scanning close to filter corners and frame interfaces.
Wide scanning footprint
The larger width transverse to the scan direction can reduce total scanning time.
Location-independent count response
Appropriate isokinetic sampling supports consistent count rates across the filter face.
Lightweight construction
Anodised-aluminium construction provides a low probe weight of approximately 0.17 kg.
Technical Specifications
| Parameter | Unit | Value / Description |
|---|---|---|
| SAMPLING | ||
| Purpose | — | Isokinetic aerosol sampling |
| Inlet flow rate | L/min | 50 |
| Applicable flow velocity | cm/s | 31–47 |
| Outlet connector | mm | Ø 10 |
| PHYSICAL | ||
| Length | mm | 150 without hose connection; 230 with hose connection |
| Material | — | Anodised aluminium |
| Weight | kg | Approximately 0.17 |
Specifications are based on Topas GmbH published product information. Confirm the current configuration, accessories and calibration scope before order placement.
Applications
Installed-filter leak testing
Scanning of installed HEPA and ULPA filter systems during leakage and integrity testing.
Isokinetic particle sampling
Representative capture of challenge aerosol downstream of filters at 50 L/min.
Cleanroom and clean-air devices
Use with suitable particle counters and scanning procedures for qualification and periodic testing.